This test is not applicable for systems that can't generate a patient print without syncing with Vitalink the vault.

The purpose of this test is to ensure that when a user alters his local scheme and creates a patient print without syncing to Vitalink the vault,
the patient can see that his patient print isn't necessarily the same as the central scheme.

...

After reading the medication scheme, the test can be executed and verified in the SUT.
A scheme will be altered using an actor without syncing to Vitalink the vault.
The patient print should be used as comparison.

...

  1. Upload the EVS export file using the EVS-action REPLACE with a patient of your choosing. The scheme should have 2 lines.
  2. Open the SUT and add a new medication to the scheme. The scheme should now have 3 lines.
  3. Generate a print without syncing to Vitalink the vault.
  • TS-1: There is a scheme with 3 lines on the patient print.
  • TS-2: There should be no reference to Vitalink the vault on the patient print.
  • TS-3: There should be no version number on the patient print.

...

  1. Upload the EVS export file using the EVS-action REPLACE with a patient of your choosing. The scheme should have 2 lines.
  2. Open the SUT and add a "dot" (.) in the description field of the line. The scheme should still have 2 lines.
  3. Generate a print without syncing to Vitalink the vault.
  • TS-1: There is a scheme with 2 lines on the patient print.
  • TS-2: There should be no reference to Vitalink the vault on the patient print.
  • TS-3: There should be no version number on the patient print.

...

  1. Upload the EVS export file using the EVS-action REPLACE with a patient of your choosing. The scheme should have 2 lines.
  2. Open the SUT and a private line to the scheme. The scheme should have 3 lines and 1 of them is private.
  3. Generate a print without syncing to Vitalink the vault.
  • TS-1: There is a scheme with 2 lines on the patient print.
  • TS-2: There should be a reference to Vitalink the vault on the patient print.
  • TS-3: There should be a version number on the patient print.
  • TS-4: The private line isn't shown on the patient print.

...

  1. Upload the EVS export file using the EVS-action REPLACE with a patient of your choosing. The scheme should have 2 lines.
  2. Open the SUT and a private line to the scheme. The scheme should still have 2 lines and 1 of them is private.
  3. Generate a print without syncing to Vitalink the vault.
  • TS-1: There is a scheme with only 1 line on the patient print.
  • TS-2: There should be no reference to Vitalink the vault on the patient print.
  • TS-3: There should be no version number on the patient print.
  • TS-4: The private line isn't shown on the patient print.

...