After reading the medication scheme, the results can be verified in the SUT.

Depending on the SUT, the handling of the medication scheme may vary slightly.


Gateway Validation Fails


Reading:

Before testing this, make sure you have uploaded the medication scheme.

Verify the following:

  • The SUT keeps functioning in a normal fashion.
  • Either there should be no medications in the UI or print, or all PLACEHOLDER AMOUNT medications are present. The SUT should not display only a set of the medicationlines.
  • The user should be able to notice that there is an abnormality with his medication scheme.


Writing:

Before testing this, make sure you have performed the 'Reading' part of the test.

Alter every medication by putting an extra dot (.) in the 'Instruction For Patient'. Then sync the altered medication scheme with Vitalink.

Verify the following:

  • The SUT should have succesfully altered the medication scheme.
  • Either there should be no medications in the UI or print, or all PLACEHOLDER AMOUNT medications are present. The SUT should not display only a set of the medicationlines.
  • The user should be able to notice that there is an abnormality with his medication scheme.