Before testing this, make sure you have uploaded the medication scheme.

Verify the following:

  • TS-1: The SUT keeps functioning in a normal fashion.
  • TS-2: Either there should be no medications in the UI or print, or all 24 medications are present. The SUT should not display only a set of the medicationlines.
  • TS-3: The user should be able to notice that there is an abnormality with his medication scheme.

...

Alter every medication by putting an extra dot (.) in the 'Instruction For Patient'. Then sync the altered medication scheme with Vitalink.

Verify the following:

  • TS-1: The SUT should have succesfully altered the medication scheme.
  • TS-2: Either there should be no medications in the UI or print, or all 24 medications are present. The SUT should not display only a set of the medicationlines.
  • TS-3: The user should be able to notice that there is an abnormality with his medication scheme.